Direct determination of Ca, Mg and Fe in serum by using a simultaneous microwave plasma torch spectrometer coupled with an improved electrothermal vaporization device

  • Jie Jiang
  • , Qiang Fei
  • , Guo Dong Feng
  • , Wen Chao Dou
  • , Ming Li
  • , Qin Han Jin
  • , Wei Jin
  • Jilin University

Research output: Contribution to journal › Article › peer-review

4 Scopus citations

Abstract

A novel procedure for the direct determination of Ca, Mg and Fe in serum using an improved electrothermal vaporization simultaneous microwave plasma torch spectrometer (ETV-MPT) was described. The background emission and spectral interferences caused by tantalum ablation were reduced by coating the coil with tantalum carbide. The proposed method eliminated the need for a lengthy sample digestion procedure, and avoided the matrix interference. The operating parameters were studied and optimized. The detection limits were 0. 04, 0.4 and 4. 6 Mu;g/L for Ca, Mg and Fe, respectively, and the relative standard deviations were less than 5 % ( n = 6). The new system was successfully applied to the analysis of practical serum sample.

Original languageEnglish
Pages (from-to)91-94
Number of pages4
JournalFenxi Huaxue/ Chinese Journal of Analytical Chemistry
Volume35
Issue number1
StatePublished - Jan 2007
Externally publishedYes

Keywords

  • Electrothermal vaporization
  • Microelements
  • Serum
  • Simultaneous microwave plasma torch-atomic emission spectrometry

Fingerprint

Dive into the research topics of 'Direct determination of Ca, Mg and Fe in serum by using a simultaneous microwave plasma torch spectrometer coupled with an improved electrothermal vaporization device'. Together they form a unique fingerprint.

Cite this