Abstract
A novel procedure for the direct determination of Ca, Mg and Fe in serum using an improved electrothermal vaporization simultaneous microwave plasma torch spectrometer (ETV-MPT) was described. The background emission and spectral interferences caused by tantalum ablation were reduced by coating the coil with tantalum carbide. The proposed method eliminated the need for a lengthy sample digestion procedure, and avoided the matrix interference. The operating parameters were studied and optimized. The detection limits were 0. 04, 0.4 and 4. 6 Mu;g/L for Ca, Mg and Fe, respectively, and the relative standard deviations were less than 5 % ( n = 6). The new system was successfully applied to the analysis of practical serum sample.
| Original language | English |
|---|---|
| Pages (from-to) | 91-94 |
| Number of pages | 4 |
| Journal | Fenxi Huaxue/ Chinese Journal of Analytical Chemistry |
| Volume | 35 |
| Issue number | 1 |
| State | Published - Jan 2007 |
| Externally published | Yes |
Keywords
- Electrothermal vaporization
- Microelements
- Serum
- Simultaneous microwave plasma torch-atomic emission spectrometry
Fingerprint