Skip to main navigation Skip to search Skip to main content

Development of a 3 T conduction-cooled animal MRI superconducting magnet with a 300 mm Warm Bore

  • Huang Xie
  • , Shunzhong Chen
  • , Wanshuo Sun
  • , Qiuliang Wang
  • Ganjiang Innovation Academy
  • CAS - Institute of Electrical Engineering

Research output: Contribution to journal › Article › peer-review

3 Scopus citations

Abstract

A 3T conduction-cooled animal MRI superconducting magnet was designed, fabricated, and tested. The MRI magnet has a large clear warm bore of Ø300 mm and a center magnetic field strength of 3 T with the measured peak-to-peak homogeneity of 217 ppm in a 180 mm diameter sphere. The magnet consists of 8 coaxial coils placed symmetrically along the axial direction. All coils made with NbTi/Cu superconducting wires were connected in series and were operated with a current of 110 A with a stored energy of about 0.62 MJ. Passive protection based on a subdivided scheme and heater network design is adopted to protect the magnet from damage caused by a quench. A two-stage GM cryocooler with a second-stage cooling power of 1.8 W is used to cool the 3 T animal MRI magnet from room temperature to 4.2 K. In this paper, magnetic design, quench protection, fabrication, and test are presented.

Original languageEnglish
Pages (from-to)1-5
Number of pages5
JournalIEEE Transactions on Applied Superconductivity
DOIs
StateAccepted/In press - 2024

Keywords

  • Coils
  • fabrication
  • Heating systems
  • Magnetic noise
  • Magnetic resonance imaging
  • Magnetic shielding
  • Magnetomechanical effects
  • MRI
  • structure design
  • superconducting magnet
  • Superconducting magnets

Fingerprint

Dive into the research topics of 'Development of a 3 T conduction-cooled animal MRI superconducting magnet with a 300 mm Warm Bore'. Together they form a unique fingerprint.

Cite this