Abstract
A flow injection chemiluminescence (CL) method is described for the determination of trenbolone acetate based on the CL generated during its reaction with KMnO 4 in acidic medium. The CL intensity is greatly enhanced by alizarin yellow R. The CL intensity is linear with trenbolone acetate concentration in the range 0.1-100.0 mg L -1 with a detection limit of 0.05 mg L -1. The sample throughout is about 90 h -1 and the relative standard deviation for 2.0 mg L -1 trenbolone acetate solution is 1.5% (n=11). The proposed method was applied to the determination of trenbolone acetate in cattle feeds.
| Original language | English |
|---|---|
| Pages (from-to) | 58-66 |
| Number of pages | 9 |
| Journal | Analytical Letters |
| Volume | 44 |
| Issue number | 1-3 |
| DOIs | |
| State | Published - Jan 2011 |
Keywords
- Cattle feed
- Chemiluminescence
- Flow injection
- Trenbolone acetate
Fingerprint
Dive into the research topics of 'Determination of trenbolone acetate in cattle feeds by a flow injection chemiluminescence method'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver