Abstract
A new method for the determination of boron by low pressure strong acid cation-exchange exclusion chromatography and the conductivity detector was proposed. The chromatographic column was filled with strong acid cation-exchange resin with improved granularity and degree of sulfonic, and mannitol-nitric acid-water was used as eluent. The method needn't a suppressor column, and the chromatographic conditions of separation and detection were modified. The detection limit was 5.0 × 10-5 mol/L, and the linear range was 0.2 × 10-3 - 5.0 × 10-3 mol/L for boron, with RSD = 2.0% (n = 11). The method was applied to the determination of boron in industrial waste water with satisfactory results.
| Original language | English |
|---|---|
| Pages (from-to) | 59-62 |
| Number of pages | 4 |
| Journal | Sichuan Daxue Xuebao (Gongcheng Kexue Ban)/Journal of Sichuan University (Engineering Science Edition) |
| Volume | 38 |
| Issue number | 1 |
| State | Published - Jan 2006 |
| Externally published | Yes |
Keywords
- Boron
- Industrial waste water
- Low pressure ion chromatography
- Retention behavior
Fingerprint