Correction of CT motion artifacts

  • Xi'an Jiaotong University
  • Air Force Medical University

Research output: Contribution to journal › Article › peer-review

Abstract

To eliminate the motion artifacts of CT images caused by patient motions and the other related errors, two kinds of correctors (A-type and U-type) are proposed to monitor the scanning process and correct the motion artifacts of the original images via reverse geometrical transformation such as reverse scaling, moving, rotating and offsetting. The results confirm that the correction method with anyone of both correctors can improve the accuracy and reliability of CT images, which facilitates eliminating or decreasing the motion artifacts, and correcting the other static errors and image processing errors, thus provides a foundation for the 3D reconstruction and the accurate fabrication of the customized implants.

Original languageEnglish
Pages (from-to)479-482
Number of pages4
JournalHsi-An Chiao Tung Ta Hsueh/Journal of Xi'an Jiaotong University
Volume38
Issue number5
StatePublished - May 2004

Keywords

  • Geometrical transformation
  • Image correction
  • Motion artifact

Fingerprint

Dive into the research topics of 'Correction of CT motion artifacts'. Together they form a unique fingerprint.