Skip to main navigation Skip to search Skip to main content

Compact radiation shield for a superconducting array molecule detector

  • Shigetomo Shiki
  • , Masahiro Ukibe
  • , Yutaka Shimizugawa
  • , Ryutaro Maeda
  • , Masataka Ohkubo
  • National Institute of Advanced Industrial Science and Technology

Research output: Contribution to journal › Article › peer-review

6 Scopus citations

Abstract

Recent development on large-scale superconducting array detectors requires a large through hole along the molecular flight path in a cryostat. The through hole causes degradation of detector performance due to the 300 K radiation and a short holding time at the cryostat base temperature. To realize a compact detector system, we designed and fabricated infrared radiation shields with a fine-honeycomb collimator and a micro-structured metal-mesh. The infrared flux through the honeycomb collimator located at 50 K was 300-500 μW. A test run in a cryostat showed a holding time of 8 hours. We fabricated a metal-mesh consisting of a self standing Cr/Cu film with an array of 2.0 μm holes having a pitch of 3.5 μm. The metal-mesh was supported by a Si reinforce structure. Spectral measurement indicates that the transmission of the 300 K radiation is less than 1%.

Original languageEnglish
Pages (from-to)495-499
Number of pages5
JournalJournal of Low Temperature Physics
Volume151
Issue number1-2 PART 1
DOIs
StatePublished - Apr 2008
Externally publishedYes

Keywords

  • Cryogenics
  • Mass spectrometry
  • Radiation shield

Fingerprint

Dive into the research topics of 'Compact radiation shield for a superconducting array molecule detector'. Together they form a unique fingerprint.

Cite this