BeO preparation and AMS measurement result for loess samples

  • Zhenkun Wu
  • , Weijian Zhou
  • , Min Liu
  • , Huagui Yu
  • , Caiwen Pan
  • , Xianghui Kong
  • , Xuefeng Lu

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

The classic procedure of BeO preparation for loess-paleosol cannot restrain the concentration of 10B effectively, which results in higher background of 10Be measurement. The procedure has been modified in this paper. It includes detailed eluting curve for cation exchanging, the method of converting Be (OH)2 to BeO, and the experiment equipments. The measurement results by Xi'an AMS indicate that the background of modified BeO preparation procedure can be lowered to 2.04 × 10-14 on 3MV AMS. The method and measurement results have also been introduced for Xi'an-AMS.

Original languageEnglish
Pages (from-to)427-432
Number of pages6
JournalHe Jishu/Nuclear Techniques
Volume31
Issue number6
StatePublished - Jun 2008