Abstract

A numerical study on the influence of the insulation layer of the hot disk sensor on the thermal conductivity measuring accuracy has been conducted. It is found that the influences of the thermal contact resistance and the insulating layer could be excluded in the transient plane source method. Both the kapton5501 and the mica5082 sensor could measure stainless steel and ceramic with a deviation less than 3% while the deviation increases to 54.2% of silica aerogel because of the large heat loss proportion through the mica5082 sensor side. The simulation proved that the heat loss through sensor side and accuracy could be improved by increasing the radius of the sensor.

Original languageEnglish
Pages (from-to)191-201
Number of pages11
JournalProgress in Computational Fluid Dynamics
Volume13
Issue number3-4
DOIs
StatePublished - 2013

Keywords

  • Accuracy
  • Experiments
  • Insulation layer influence
  • Numerical
  • Thermal conductivity
  • Transient plane source