Skip to main navigation Skip to search Skip to main content

A 0.0075 mm2, 0.0139 FoM Double-Ring Vernier TDC with Vernier Arbiter Phase Calibration and Code Error Cancellation for 26 dB SFDR Improvement in ToF-PET

  • Han Wang
  • , Youze Xin
  • , Yiyun Xie
  • , Pengfei Hu
  • , Hongxu Chen
  • , Ziyang Xu
  • , Xuhu Wang
  • , Congzhen Hu
  • , Chi Wang
  • , Lu Li
  • , Lei Fang
  • , Li Geng
  • , Bing Zhang
  • Xi'an Jiaotong University
  • Huazhong University of Science and Technology
  • Ltd.

Research output: Contribution to journal › Article › peer-review

Abstract

This work presents a PET-oriented double-ring Vernier time-to-digital converter (DRVTDC) that targets the timing-quantization block required in multi-voltage-threshold (MVT) and time-over-threshold (ToT) front-end readout for time-of-flight (ToF) positron emission tomography (PET) applications. This work addresses two critical error mechanisms that limit the use of multi-stage Vernier TDCs: parallel output misalignment (POM) between quantization stages and mismatch-induced chasing error (MCE) in the Vernier arbiter array. To suppress these errors, a joint calibration scheme combining code error cancellation (CEC) and Vernier arbiter phase calibration (VAPC) is proposed. The core TDC achieves a 10.4-ps resolution with only 0.0075 mm2 core area. The implemented 7-bit coarse counter provides a dynamic range (DR) of 5.12 μs. Empowered by these techniques, the measured DNL and INL ranges are significantly improved from +3.17/−0.89 LSB and +7.0/−6.9 LSB to +0.34/−0.29 LSB and +1.15/−1.09 LSB, respectively, alongside a 26-dB enhancement in the SFDR. The core TDC consumes 1.758 mW, with a total of 5.526 mW including the PLLs.

Original languageEnglish
JournalIEEE Transactions on Biomedical Circuits and Systems
DOIs
StateAccepted/In press - 2026

Keywords

  • code error cancellation (CEC)
  • double-ring Vernier
  • mismatch-induced chasing error (MCE)
  • parallel output misalignment (POM)
  • Positron emission tomography (PET)
  • time-to-digital converter (TDC)
  • Vernier arbiter phase calibration (VAPC)

Fingerprint

Dive into the research topics of 'A 0.0075 mm2, 0.0139 FoM Double-Ring Vernier TDC with Vernier Arbiter Phase Calibration and Code Error Cancellation for 26 dB SFDR Improvement in ToF-PET'. Together they form a unique fingerprint.

Cite this