Abstract
Objective: Aiming at the clinical problem of the low matching degree with the patient's anatomical morphology for traditional cervical fusion cage, a cervical fusion cage with the function of adjustable height and the shape matched with the vertebral body was established, and its biomechanical properties were evaluated. Methods: A cervical C4-5 segment fusion model was established according to anterior cervical discectomy and fusion (ACDF), so as to simulate different motion conditions, i.e. anterior flexion, posterior extension, left/right lateral flexion, left/right rotation, and stress of the fusion cage and vertebral endplate was calculated. After three-dimensional (3D) printing of the fusion cage, an in vitro mechanical experiment was conducted to explore safety and stability of the fusion cage. Results: The fusion cage could keep the range of motion (ROM) of cervical vertebrae at the fusion segment with 1°-2.8° and reduce the ROM to 40%-80% of the natural segment. In the in vitro compression test, the yield load of the fusion cage was (2 721.67±209) N, which met the maximum demand of the physiological load in service state. Conclusions: The designed fusion device with adjustable height shows better biomechanical properties and can reduce the selection step in operation.
| Translated title of the contribution | Optimization Design and Evaluation of Three-Dimensional Printed PEEK Cervical Interbody Fusion Cage with Adjustable Height |
|---|---|
| Original language | Chinese (Traditional) |
| Pages (from-to) | 177-183 |
| Number of pages | 7 |
| Journal | Yiyong Shengwu Lixue/Journal of Medical Biomechanics |
| Volume | 36 |
| Issue number | 2 |
| DOIs | |
| State | Published - Apr 2021 |
Fingerprint
Dive into the research topics of 'Optimization Design and Evaluation of Three-Dimensional Printed PEEK Cervical Interbody Fusion Cage with Adjustable Height'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver